Daily QA for Linacs

DOSIMETRY TOLERANCE
X-ray output constancy 3%
Electron output constancy*** 3%

MECHANICAL
Localizing lasers 2mm
Distance indicator (ODI) 2mm

SAFETY
Door interlock Functional
Audiovisual monitor Functional


**** All electron energies need not be checked daily, but all electron energies are to be checked at least twice weekly.